Infant incubator
for premature infants, low birth weight infants, critically ill
children, newborn to provide a similar training of maternal uterine
environment equipment, temperature can also be used for the recovery of
the infant, infusion, rescue, and so on hospitalized for observation.
As the infant's own ability to resist very weak, which requires a
stable incubator performance, can provide a more closer to the baby in
the mother's womb environment more conducive to the health of babies.
Health care workers to understand the basic working principle of their
operation to the correct use of equipment performance can be played
more.
1, the basic principles of infant incubator
Incubator production currently on the market there are many
manufacturers, the basic principles of more or less the same, most have
a "convective thermal regulation" provided in the form of an air
purification, temperature and humidity environment of a quality
suitable for the use of computer technology on the incubator
temperature (Box Temperature / skin temperature) to implement the servo
control, boot automatically into the state of temperature control box.
Temperature controller is the core component of the equipment,
temperature settings, real-time temperature monitoring and other
functions, when the medical staff can set the temperature when the
temperature inside the continuous surveillance. Machinery required to
open the perspective of operation and from time to time to change the
crib Shizuma good results, so that medical personnel in the care of
rest will not affect the baby or the baby woke up, effectively reducing
the infant's energy metabolism.
Incubators need to have a reliable alarm system failure, to achieve
power, box temperature, skin temperature and the flow duct of the
monitoring, if any of the indicators should have exceeded the allowed
range of sound and light alarm, prompt attention to the guardianship
staff. Incubator alarm when the medical staff found to be an immediate
end to the cultivation of the infant, the infant moved to a safe place
to ensure the safety of the baby and immediately notify the
professional maintenance engineer to carry out maintenance on the
equipment, only when the failure to continue to use ruled out entirely.
Incubator use of AC220V power source, must be in strict compliance with
the single-phase three-wire system of principles, to ensure a good
grounding, apparatus with a network interface for the general infant
care network with the infant incubator system to provide a dedicated
outlet.
2, the safe use of
Infant Incubator
with the work of the state of relations between the baby's safety in
the use of health care does not want to be very careful. Before use, to
deal with the various functions of the system of checks seriously.
(1) of the power failure, the inspection report to the police function:
training course in infants who have any power, would lead to dramatic
changes in temperature, endangering the baby's life. In the use of this
feature before the inspection, disconnect the power supply AC220V
incubator and connected to temperature controller opens the power
switch off after the alarm devices should appear, turn off the power
switch after the alarm will automatically eliminate. Alarm power supply
is provided by the battery, under normal use of the system will
automatically charge the battery, the battery has a certain period of
use to the period should be to replace the battery.
(2) inspection of the temperature controller: Temperature Controller
through the sensor, fan and heating pipe to complete the baby inside of
the heating and air circulation, regardless of which components will
directly affect the damage to the environment in which the baby. AC220V
power supply connected, open the power switch humidity controller,
temperature indicator light settings box, set the temperature display
flashes to show the temperature control box set the default value 32 ?,
do not make any operation, apparatus, automatic temperature control to
enter the boxes, and real-time temperature show me the window display
of real-time temperature.
(3) over-temperature alarm function checks a set of incubator are
generally not subject to the temperature controller to control the
independent over-temperature alarm system, temperature control
instrument for the failure to monitor the temperature of the box, a
measure that greatly the incubator to raise the safety and reliability.
Inspection, in boxes under temperature control, set the value to be set
at 36 ?, after entering the thermostat, quickly followed by Canada and
by key, when the set temperature display window does not show that the
temperature controller to enter a state of over-temperature experiment,
this heating power when the lights all lit. After about 10min,
apparatus should appear over-temperature alarm. By the reset button
after the end of over-temperature alarm, apparatus to return to work.
(4) see the inspection report to the police machine: machines work,
hold down the finger "to stop fans to" test button, so that fan speed
to reduce or stop the rotation, the apparatus should appear bright
lights and fan alarm sound alerts.
Confirmed by the inspection device are normal functions, in accordance
with the orders set up the parameters to allow the machine to enter the
pre-state of the heating apparatus generally in about 45min, but the
warm-up time normally required 2h to reach the temperature inside the
balance infants are allowed to carry out the training. Show real-time
temperature in the instrument settings to achieve the start of the baby
on the train is not correct.
Upon completion of the training for each infant, the incubator must be
a thorough cleaning, sterilization, disinfection (using the process at
least once a week), for some components to be demolished by the
cleaning.
The relationship between
infant incubator to culture baby's effectiveness and safety in the use of standards when we must, to prevent the accident.